Home / Early Warning / Vulnerabilidades / CVE-2015-3956

CVE-2015-3956

Type: 
Insufficient Verification of Data Authenticity
Severity: 
High
Publication date: 
03/25/2019
Last modified: 
10/09/2019
Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Impact
Access Vector: Through network
Access Complexity: Low
Authentication: Not required to exploit
Impact Type: Total commitment on system integrity + Total commitment on system confidentiality + Total commitment on system availability
Vulnerable software and versions
  • cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*
  • cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:*
  • cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:*
  • cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*
  • cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:*
  • cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:*
To consult the complete list of products and versions see this page
References to Advisories, Solutions, and Tools